Accuracy of an Endodontic Rotary Motor with Integrated Apex Locator to Determine the Root Canal Length

Authors

  • Borges, JHS Universidade Federal da Bahia Author
  • Carvalho FB Universidade Federal da Bahia Author
  • Cangussu MCT Universidade Federal da Bahia Author
  • Peters OA The University of Queensland
  • Bahia MGA Universidade Federal de Minas Gerais Author
  • Érika Sales Joviano Pereira Universidade Federal da Bahia Author

DOI:

https://doi.org/10.35699/2178-1990.2023.45739

Keywords:

Endodontia, Cavidade pulpar dentária, Pesos e medidas

Abstract

Aim: The purpose of this study was to evaluate the efficacy to determine the root canal length, in vitro, of both the electronic apex locator (M2) and the autostop (AS - M3) functions of the Endus Duo Gnatus endodontic motor (Gnatus, São Paulo, SP, Brazil).

Methods: Thirty extracted human single-rooted premolars had their root canal lengths (TLs) up to the apical foramen determined using the Endus Duo Gnatus in two ways: (1) In the stainless steel (SS) control group, the measurement was obtained using a stainless steel hand instrument with the electronic locator mode (M2 function) connected to a stainless-steel hand instrument (K-file #15). In the NiTi rotary instrument (NiTi RI) intervention group, the measurement was obtained during the instrumentation (M3 function) of the root canals with a nickel-titanium rotary instrument (Hyflex CM: Coltene Whaledent, Cuyahoga Falls, OH, USA), size 25/.06. The NiTi manually used instrument (NiTi MUI) Intervention group performed the measurement in locator mode using a Hyflex instrument, placed to true length manually. Statistical analysis was performed using ANOVA followed by the Tukey post-hoc test with a significance level of p < 0.05.

Results: The true mean length and standard deviation (SD) of the standardized root canals were 18.40 ± 2.14mm, while the mean lengths and standard deviations (SD) were 18.29 ± 1.89mm, 18.22 ± 1.85mm, and 17.24 ± 2.09mm for the SS, NiTi RI, and NiTi MUI groups, respectively. However, data from the NiTi MUI Intervention group indicated shorter root canal lengths when compared to the SS control group and the NiTi RI Intervention group values, and were significantly shorter than the true canal length (p < 0.001).

Conclusions: The use of the motor in NiTi RI Intervention group showed acceptable results. However, the NiTi MUI Intervention group resulted in unacceptable short measurements.

 

References

Ricucci D, Langeland K. Apical limit of root canal instrumentation and obturation, part 2. A histological study. Int Endod J. 1998;31(6):394–409.

Sjögren U, Hägglund B, Sundqvist G, Wing K. Factors affecting the long-term results of endodontic treatment. J Endod. 1990;16(10):498–504.

Cruz ATG, Wichnieski C, Carneiro E, Neto UXS, Gambarini G, Piasecki L. Accuracy of 2 endodontic rotary motors with integrated apex locator. J Endod. 2017;43(10):1716-9.

Kuttler Y. Microscopic investigation of root apexes. J Indiana Dent Assoc. 1955;89(1):20-8.

Wu MK, Wesselink PR, Walton RE. Apical terminus location of root canal treatment procedures. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2000;89(1):99–103.

Gordon M, Chandler N. Electronic apex locators. Int Endod J. 2004;37(7):425–37.

Dummer PM, McGinn J, Rees DG. The position and topography of the apical canal constriction and apical foramen. Int Endod J. 1984;17(4):192–8.

Elayouti A, Hülber-J M, Judenhofer MS, et al. Apical constriction: location and dimensions in molars - a micro–computed tomography study. J Endod. 2014;40(8):1095–9.

Nekoofar M, Ghandi M, Hayes S, Dummer P. The fundamental operating principles of electronic root canal length measurement devices. Int Endod J. 2006;39(8):595–609.

Plotino G, Grande N, Brigante L, et al. Ex vivo accuracy of three electronic apex locators: Root ZX, Elements Diagnostic Unit and Apex Locator and ProPex. Int Endod J. 2006;39(12):408–14.

Piasecki L, Carneiro E, Silva Neto UX, et al. The use of micro–computed tomography to determine the accuracy of 2 electronic apex locators and anatomic variations affecting their precision. J Endod. 2016;42(8):1263–7.

Martins JN, Marques D, Mata A, Caramês J. Clinical efficacy of electronic apex locators: systematic review. J Endod. 2014;40(6):759–77.

Vasconcelos BC, Bastos LM, Oliveira AS, et al. Changes in root canal length determined during mechanical preparation stages and their relationship with the accuracy of Root ZX II. J Endod. 2016;42(11):1683–6.

Fadel G, Piasecki L, Westphalen V, et al. An in vivo evaluation of the auto apical reverse function of the Root ZX II. Int Endod J. 2012;45(10):950–4.

Vasconcelos BC, Frota LM, Abreu Souza T, et al. Evaluation of the maintenance of the apical limit during instrumentation with hybrid equipment in rotary and reciprocating modes. J Endod. 2015;41(5):682–5.

Siu C, Marshall JG, Baumgartner JC. An in vivo comparison of the Root ZX II, the Apex NRG XFR, and Mini Apex Locator by using rotary nickel-titanium files. J Endod. 2009;35(7):962–5.

Carneiro E, Bramante CM, Picoli F, et al. Accuracy of root length determination using Tri Auto ZX and ProTaper instruments: an in vitro study. J Endod. 2006;32(2):142–4.

Felippe W, Felippe M, Reyes Carmona J, et al. Ex vivo evaluation of the ability of the Root ZX II to locate the apical foramen and to control the apical extent of rotary canal instrumentation. Int Endod J. 2008;41(6):502–7.

Jakobson SJ, Westphalen VP, Silva Neto UX, et al. The accuracy in the control of the apical extent of rotary canal instrumentation using Root ZX II and ProTaper in- struments: an in vivo study. J Endod. 2008;34(11):1342–5.

Akisue E, Gratieri SD, Barletta FB, Caldeira CL, Grazziotin-Soares R, Gavini G. Not all electronic foramen locators are accurate in teeth with enlarged apical foramina: an in vitro comparison of 5 brands. J Endod. 2014;40(1):109-12.

Goldberg F, Silvio AC, Manfre S, Nastri N. In vitro measurement accuracy of an electronic apex locator in teeth with simulated apical root resorption. J Endod. 2002;28(6):461-3.

ElAyouti A, Dima E, Ohmer J, et al. Consistency of apex locator function: a clinical study. J Endod. 2009;35(2):179–81.