Effective dose and detriment assessment in cone beam computed tomography (CBCT)

Authors

  • Virgínia Medeiros Ferreira Universidade Federal de Minas Gerais
  • Keli Bahia Felicíssimo Zocratto Centro Universitário Newton Paiva
  • Cláudia Borges Brasileiro Universidade Federal de Minas Gerais

DOI:

https://doi.org/10.7308/aodontol/2012.48.4.07

Keywords:

Cone beam computed tomography, Radiation dosage, Risk assessment

Abstract

Aim: To calculate the effective dose (E) and detriment associated with cone beam computed tomography (CBCT), following recommendations set forth by the International Comission on Radiological Protection (ICRP).

Materials and Methods: The effective dose (E) was calculated by multiplying the absorbed/equivalent doses by the bone marrow, thyroid, skin, bone surface, salivary glands, brain, and remaining organs (extrathoracic region, lymphatic nodes, muscles, and oral mucosa), resulting from different equipments and imaging protocols, by tissue weighting factor of these tissues. Detriment was estimated by multiplying E by the probability coefficient for stochastic effects after exposure to low doses of radiation: 5.7 x 10-2 Sv-1. The equipment and protocols were grouped according to the irradiated volume (field of view - FOV). The mean effective doses, standard deviation, and variation coefficient of each group were also calculated.

Results: Protocols that employed a larger FOV resulted in an increase of the effective dose and detriment. The mean (standard deviation) of E for small, medium, and large FOV were 33.91 µSv (13.38), 82.85 µSv (74.26), and 107.99 µSv (43.82), respectively. The small FOV group presented the lowest variation coefficient (39%).

Conclusion: The E from CBCT are high when compared to other conventional radiography, although the detriment is relatively low. Recommendations for CBCT and the choice of protocol must be carefully justified so that the benefits of patient exposure outweigh the potential radiation detriment.

Downloads

Download data is not yet available.

References

Ludlow JB, Davies-Ludlow LE, White SC. Patient risk related to common dental radiographic examinations: the impact of 2007 International Commission on Radiological Protection regarding dose calculation. J Am Dent Assoc. 2008; 139:1237-43.

Garib DG, Raymundo Jr R, Raymundo MV, Raymundo DV, Ferreira SN. Tomografia computadorizada de feixe cônico (cone beam): entendendo este novo método de diagnóstico por imagem com promissora aplicabilidade na ortodontia. Rev Dental Press Ortodon Ortop Facial. 2007; 12:139-56.

Mohan R, Singh A, Gundappa M. Three dimensional imaging in periodontal diagnosis: utilization of cone beam computed tomography. J Indian Soc Periodontol. 2011; 15:11-7.

Roberts JA, Drage NA, Davies J, Thomas DW. Effective dose from cone beam CT examinations in dentistry. Br J Radiol. 2009; 82:35-40.

Qu X-M, Li G, Ludlow JB, Zhang Z-Y, Ma X-C. Effective radiation dose of ProMax 3D cone-beam computerized tomography scanner with different dental protocols. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2010; 110:770-6.

Torres MGG, Campos PSF, Neto Segundo NP, Ribeiro M, Navarro M, Crusoé-Rebello I. Avaliação de doses referenciais obtidas com exames de tomografia computadorizada de feixe cônico adquiridos com diferentes tamanhos de voxels. Dental Press J Orthod. 2010; 15:42-3.

Pauwels R, Beinsberger J, Collaert B, Theodorakou C, Rogers J, Walker A, et al. Effective dose range for dental cone beam computed tomography scanners. Eur J Radiol. 2012; 81:267-71.

Wrixon AD. New ICRP recommendations. J Radiol Prot. 2008; 28:161-8.

Ludlow JB, Ivanovic M. Comparative dosimetry of dental CBCT devices and 64-slice CT for oral and maxillofacial radiology. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2008; 106:106- 14.

Wrixon AD. New recommendations from the International Commision on Radiological Protection: a review. Phys Med Biol. 2008; 53:R41-R60.

International Commission on Radiological Protection. The 2007 Recommendations of the International Commission on Radiological Protection. ICRP Publication 103. Ann ICRP. 2007; 37:1-332.

Gibbs SJ, Pujol Jr. A, Chen T-S, James Jr. A. Patient risk from intraoral dental radiography. Dentomaxillofac Radiol. 1988; 17:15-23.

Hatcher, D. Operational principles for cone-beam computed tomography. J Am Dent Assoc. 2010; 141:3S-6S.

Scarfe WC, Farman AG. What is cone-beam CT and how does it work. Dent Clin N Am. 2008; 52:707-30.

Tetradis S, Anstey P, Graff-Radford S. Cone beam computed tomography in the diagnosis of dental disease. J Calif Dent Assoc. 2010; 138:27- 32.

Loubele M, Bogaerts R, Van Dijck E, Pauwels R, Vanhersden S, Suetens P, et al. Comparison between effective radiation dose of CBCT and MSCT scanners for dentomaxillofacial applications. Eur J Radiol. 2009; 71:461-8.

Published

2016-06-10